2.2 X-ray interface box
The X-ray interface box monitors the statuses of the X-ray generator and detector to control the
X-ray exposure and imaging. It consists of the PCA-XIF board, AC adapter, X-ray interface
cable, grounding wire and covers. Even when a multiple number of detectors are in use, only
one detector per X-ray interface box can be specified from the image capture computer to take
images.
Image
Capture
Computer
Detector
X-ray
interface box
X-ray generator
X-ray exposure
Exposure switch
TCP/IP connection
Detector selection
Imaging condition
setting
Imaging data transfer
Hub/wireless AP
Wi
rele
ss
L
A
N
*
Wired LAN
*Either a wireless or wired format can be used for the LAN connection of the detector.
In the imaging sequence described below, the items printed in bold represent the control
operations performed by the X-ray interface box.
(1) The user presses the exposure switch.
(2) The
exposure enable request signal
from the X-ray generator is
monitored by the X-ray
interface box
.
(3) The X-ray interface box makes an
Enable judgment
for the
exposure enable request
signal
, and
sends the trigger signal
to the detector.
(4) When the imaging preparations of the detector are completed, the imaging enable signal is
sent from the detector to the X-ray interface box.
(5) The
X-ray interface box monitors the imaging enable signal
from the detector.
(6) The X-ray interface box
makes an Enable judgment for the imaging enable signal
, and
sends the exposure enable signal
to the X-ray generator.
(7) With the detector now in the imaging status, X-ray exposure is commenced.
(8) When the detector has finished imaging, the images are transferred to the image capture
computer.
(9) This completes the imaging sequence.