system for the installation site
2) Power consumption [W]: Information for the purpose of designing the power supply
system for the installation site (Note that this is valid only
when the power factor is high [ex. 90% or more].)
3) Heat generation [cal or J]: Information for the purpose of designing the cooling system
for the installation site
*2
: The heat generation is calculated from the power consumption (W) as follows:
Heat generation (kcal/h) = power consumption (W) * 0.23889 * 60 * 60 / 1000
Heat generation (kJ/h) = power consumption (W) * 60 * 60 / 1000