2. Installation procedures
- 28 -
5) Upon completion of the connections, attach the covers of the X-ray interface box.
* Take care not to strip the screw heads of the side panel and bottom panel.
* The screws used for mounting must be tightened up in the following sequence: bottom panel
(screws with external teeth)
→
both side panels (brazier head screws).
6) Connect the AC power cable to the AC adapter.
7) Connect the supplied grounding wire (functional ground)
to the ground terminal on the X-ray
interface box. The grounding wire is used for functional ground, but not protective earth. To prevent the
equipments from being damaged, connect to other than potential equalization bus (grounding wire of the
distribution panel or iron frame)
8) Connect the LAN cable and DC plug (of the AC adapter) to the connector connection area of the
X-ray interface box.
* Be absolutely sure to use the AC adapter and AC power cable which come supplied as
accessories with this product.
* The supplied ferrite core must be installed by looping the LAN cable around it at a position
not more than 10 cm (shown by the shading in the figure below) from the connector of the
LAN cable.
AC adapter
DC plug
AC power cable
Power LED
DC plug
X-ray interface cable
Grounding wire
Less than 10cm
Ferrite core
Grounding wire
LAN cable