2. Installation procedures
- 33 -
7.1.10 When using an existing detector together with the CXDI-80C Wireless System
CXDI-80C SYSTEM and other Ferry supported detector systems (CXDI-70C Wireless,
401G/401C, 401G COMPACT/401C COMPACT, 501G/501C), and the existing CXDI
SYSTEM (CXDI-40EG/40EC, 50G/50C, 40G COMPACT, 55G/55C, 60G/60C) can be used
together. In a case like this, follow the steps below for installation.
[Setup flow: Combining the existing CXDI CXDI-80C SYSTEM]
(1) Configure the existing CXDI SYSTEM.
(2) Connect the existing CXDI SYSTEM to Image Capture Computer through the switching
hub.
(3) Turn on the power to the Image Capture Computer and install CXDI Control Software NE
Ver.1.4 or later.
(4) Start up the service tool, and press the “ENROLL” button.
(5) Check that the detector serial number of the existing detector is displayed on the Connection
tab. For details, refer to the “CXDI Control Software NE Installation Manual.”
(6) Configure CXDI-80C SYSTEM and perform LINK for sensor recognition.
(7) Check that the detector serial number of 80C Wireless is displayed on the Connection tab.
[Setup flow: Combining CXDI-80C CXDI-70C detector]
(1) Configure the CXDI-80C SYSTEM.
(2) Connect CXDI-80C SYSTEM to the Image Capture Computer through the switching HUB.
(3) Turn on the power to the Image Capture Computer, and install CXDI Control Software NE
Ver.1.4 or later.
(4) Recognize the 80C detector with the sensor using infrared ray communication.
(5) Check that the detector serial number for the 80C detector is displayed on the Connection
tab.
(6) Recognize the 70C detector with the sensor using infrared ray communication.
(7) Check that the detector serial number for the 70C detector is displayed on the Connection
tab.
[Setup flow: Combining CXDI-80C other Ferry supported detector systems]
(1) Configure another Ferry supported detector(Example: CXDI-401G SYSTEM).
(2) Connect CXDI-401C SYSTEM to the Image Capture Computer through the switching HUB.
(3) Turn on the power to the Image Capture Computer, and install CXDI Control Software NE
Ver.1.4 or later.
(4) Startup the service tool, then press [ENROLL].
(5) Check that the detector serial number for the 401C detector is displayed on the Connection
tab.
(6) Configure the CXDI-80C SYSTEM.
(7) Connect CXDI-80C SYSTEM to the Image Capture Computer through the switching HUB.
(8) Recognize the 80C detector with the sensor using infrared ray communication.
(9) Check that the detector serial number for the 80C detector is displayed on the Connection
tab.
Summary of Contents for CXDI-80C
Page 6: ...Note These measurements are only guidelines ...
Page 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Page 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Page 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...