6. Troubleshooting
- 13 -
No.5
If Collection Tool is started up and the [Other Information] category is selected, the screwcap.log,
cxdcap.log and other log files can be collected.
No.6
If Collection Tool is started up and the [Other Information] category is selected, [SnsPBInfo] is
generated automatically, and the DP files and detector logs and X-ray interface box logs can be
collected.
No.7
If Collection Tool is started up and the [Other Information] category is selected, [elog] is
generated automatically, and the following four event viewer logs can be collected.
C:¥Windows¥System32¥winevt¥Logs
•
Application
•
Security
•
Setup
•
System
[Manual intervention required for some logs]
1)
Logs of X-ray interface box (PCA-XIF board)
•
Log collection is required when the X-ray communication failure and/or the system error occurs. For
details on the procedures, please refer to 5.3.2 Obtaining the PCA-XIF board log in Chapter 4 Repair
guide.
•
Specify the SnsPBInfo folder as a save folder of the logs.
2)
Sensor logs
•
Sensor logs must be required when the system error, wireless communication failure and/or image
problem occurs.
•
Start Ferry Type FPD Maintenance Tool, select Target IP Address of the detector of which logs you
are going to collect and press the Read Log button.
•
Sensor logs are saved in the SnsPBInfo folder.
•
For details on the procedures, please refer to CXDI Control Software NE Maintenance Manual.
To collect the logs of X-ray interface box and sensor logs by the Collection Tool, be sure to collect
them and save in the SnsPBInfo folder manually.
Summary of Contents for CXDI-80C
Page 6: ...Note These measurements are only guidelines ...
Page 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Page 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Page 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...