3
3
Description of Functions
- 2 -
1.2 Loading and outputting the data of the images taken
• When viewed from front panel of detector
(1) The row of the data to be output is selected by Drv-IC.
(2) The data to be output from the row selected by Drv-IC is loaded by Amp-IC, A/D
conversion is performed, and then the data is output to the PCA-FE board.
②
①
Drv-IC
0LB11 0LB10 0LB9 0LB8
0LB7
0LB6
0LB5
0LB4 0LB3
0LB2 0LB1
0LB1
0L
B2
0L
B3
0L
B4
0L
B5
0L
B6
0L
B7
0L
B8
0L
B9
ADCLK2
AD_DATA_B
AD_DATA_A
ADCLK1
PCB-A
PCB-
D
21
92
1528
1272
2800
: System home point
Direction of X-ray
incidence
Amp-IC