2. Installation procedures
-
43
-
7.5 Image quality checks
(1) Objective
To check the final quality of the images.
(2) Tools used
Radiographic phantom
<Procedure>
Step/item Description/conditions
1.
Distance between X-ray bulb focal point and detector
Xcm: Based on grid used
2.
Proceed with calibration.
• The calibration must be completed successfully.
3.
Proceed with Self-diagnosis and Performance test.
• All the items must pass the self-diagnosis.
X-ray tube
focal point
Xcm
Center line at
right angle
CXDI detector
Self-diagnosis
<Positioning>
Align the positions between the
X-ray tube and detector.
Calibration