38
of 148
Display of the
measurement parameter menu
(page 3)
INPUT
HOME
RETU
A/D CONV:
STANDARD
MEAS. -SQ.:
SINGLE COMP.
MEAS. PARAMETER S 3
Various measurement parameters can be set in this menu.
The following entries are permissible:
The [F3] key (HOME) effects a return to the main menu.
The [F4] key (RETUrn) effects a return to the parameter selection menu.
This menu display covers 3 pages.
The [
▲
] and [
▼
] keys are used to switch between the three pages.
The [
↑
] and [
↓
] cursor keys are used to select the required parameter on the current page. The selected
parameter is displayed inversely.
The [POS1] key is used to select the 1st parameter in a window.
The [END] key is used to select the last parameter in a window.
Specifying the
SELECT RANGE
The [
→
] and [
←
] cursor keys are used to select one of the following options:
MAN: The measuring range must be selected manually.
AUTO: The RESISTOMAT® model 2329 automatically selects the appropriate measuring range.
On a change from MAN to AUTO, the RESISTOMAT® model 2329 first selects the 200 kOHM range.
A change between MAN and AUTO can also be made with the A/M key in the main menu.
The following display appears in the AUTO mode:
A-HI
HOME
RETU
MEAS. PARAMETER S 1
SELECT RANGR:
AUTO
MIN. RGE:
200 mOHM
LOAD:
R
MEAS. MODE:
SINGLE
Display of the
measurement parameter menu
(page 1)
when AUTO is selected
When automatic range selection is ac-
tive, the permissible measuring ranges
can be limited in order to optimize the
measurement rate.
For example: A - HI = 200
Ω
A - LO = 2
Ω
With these settings, the RESISTOMAT®
model 2329 only measures in the 2
Ω
,
20
Ω
and 200
Ω
ranges.
The currently smallest measuring range
is indicated.