124
of 148
DA? (short form)
DESCRIPTION:
This retrieves 20 measurement values stored in a datalogger block.
SYNTAX:
DA? <bl>, <num>
PARAMETERS:
bl: Block number 0 to 31
num: Number of the measurement value in the block (num >=1)
QUERY FORM:
Only query form
RESPONSE:
20 consecutive measurement values, starting with the value at memory location
<num>.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DATalogger:DATA:DATE?
DESCRIPTION:
This retrieves the storage date of a measurement value.
SYNTAX:
DATalogger:DATA:DATE? <bl>, <num>
PARAMETERS:
bl: Block number 0 to 31
num: Number of the measurement value in the block
(num >=1)
QUERY FORM:
Only query form
RESPONSE:
The storage date of the selected resistance value (dd, mm, yyyy)
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DATalogger:DATA:TIME?
DESCRIPTION:
This retrieves the storage time of a measurement value.
SYNTAX:
DATalogger:DATA:TIME? <bl>, <num>
PARAMETERS:
bl: Block number 0 to 31
num: Number of the measurement value in the block (num >=1)
QUERY FORM:
Only query form
RESPONSE:
The storage time of the selected resistance value.
(hh, mm, ss, xx) xx -> 1/100 seconds
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.