125
of 148
DATalogger:DATA:ALL?
DESCRIPTION:
This retrieves a stored measurement value together with the corrseponding
date and time.
SYNTAX:
DATalogger:DATA:ALL? <bl>, <num>
PARAMETERS:
bl: Block number 0 to 31
num: Number of the measurement value in the block (num >=1)
QUERY FORM:
Only query form
RESPONSE:
The stored measurement value together with the
corrseponding date and time.
Measurement value, date, time.
Meas. value: Resistance value with unit
Date: dd, mm, yyyy
Time: hh, mm, ss, xx (xx -> 1/100 seconds)
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DV? (short form)
DESCRIPTION:
This retrieves 7 stored measurement values together with the corrseponding
dates and times.
SYNTAX:
DV? <bl>, <num>
PARAMETERS:
bl: Block number 0 to 31
num: Number of the measurement value in the block (num >=1)
QUERY FORM:
Only query form
RESPONSE:
7 consecutive measurement values together with the
corrseponding dates and times of storage.
The first value refers to the memory location.
Measurement value, date, time.
Meas. value: Resistance value with unit
Date: dd, mm, yyyy
Time: hh, mm, ss, xx (xx -> 1/100 seconds)
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DATalogger:SIZE
DESCRIPTION:
This specifies the size of a datalogger block.
SYNTAX:
DATalogger:SIZE <bl>, <size>
PARAMETERS:
bl: Block number 0 to 31
size: Size of the datalogger block
QUERY FORM:
DATalogger:SIZE? <bl>
RESPONSE:
The size of the datalogger block with the number <bl>.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.