112
of 148
6.3.13 ACCess Subsystem
ACCess:MEASure
DESCRIPTION:
This enables/disables access to the measurement-parameter menu.
SYNTAX:
ACCess : MEASure <parameter>
PARAMETERS:
1 or ON -> Enables access to the measurement-parameter menu.
0 or OFF ->Disables access to the measurement-parameter menu.
QUERY FORM:
ACCess:MEASure?
RESPONSE:
1 ->If access to the measurement-parameter menu is enabled.
0 ->If access to the measurement-parameter menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
ACCess:EXPRession
DESCRIPTION:
This enables/disables access to the display-mode menu.
SYNTAX:
ACCess : EXPRession <parameter>
PARAMETERS:
1 or ON -> Enables access to the display-mode menu.
0 or OFF ->Disables access to the display-mode menu.
QUERY FORM:
ACCess:EXPression?
RESPONSE:
1 ->If access to the display-mode menu is enabled.
0 ->If access to the display-mode menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
ACCess:PRINter
DESCRIPTION:
This enables/disables access to the printer menu.
SYNTAX:
ACCess:PRINter <parameter>
PARAMETERS:
1 or ON -> Enables access to the printer menu.
0 or OFF -> Disables access to the printer menu.
QUERY FORM:
ACCess:PRINter?
RESPONSE:
1 -> If access to the printer menu is enabled.
0 ->If access to the printer menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.