102
of 148
CALCulate:LIMit:ACKNowledge?
DESCRIPTION:.
This tests and saves the comparator limits.
SYNTAX:
CALCulate:LIMit:ACKNowledge?
PARAMETERS:
None
QUERY FORM:
Only query form
RESPONSE:
1 -> If the limit test is OK, the values are saved.
0 -> If the limit test is not OK, the values are not saved.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
CALCulate:LIMit:CLEar
DESCRIPTION:
This deletes the statistical results provided by the comparator.
SYNTAX:
CALCulate:LIMit:CLEar
PARAMETERS:
None
QUERY FORM:
No query form
RESPONSE:
—
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
CALCulate:LIMit:REPort
DESCRIPTION:
This requests the statistical results provided by the comparator.
SYNTAX:
CALCulate:LIMit:REPort?
PARAMETERS:
None
QUERY FORM:
Only query form
RESPONSE:
If 2 limits are set: value1, value2, value3
value1: No. of meas. values < lower limit
value2: Lower limit >=No. of meas. values<=upper limit
value3: No. of meas. values > upper limit
If 4 limits are set: value1, value2, value3, value4, value5
value1: No. of meas. values < limit1
value2: limit1 >= No. of meas. values < limit2
value3: limit2 >= No. of meas. values <= limit3
value4: limit3 > No. of meas. values <= limit4
value5: No. of meas. values > limit4.
RESTRICTIONS: