99
of 148
CALCulate:LIMit:LOWer
DESCRIPTION:
This specifies the lower limit when 2 comparator limits are selected.
SYNTAX:
CALCulate:LIMit:LOWer <parameter>
PARAMETERS:
The lower limit with an optional unit. Valid units: UOHM, MOHM,
OHM, KOHM, MAOHM. If no unit is specified, the unit of the value is
interpreted as OHM.
QUERY FORM:
CALCulate:LIMit:LOWer?
RESPONSE:
The lower limit of 2 comparator limits (value with unit).
BEMERKUNGEN:
The value transferred with this command needs to be acknowledged
with CALCulate:LIMit:ACKNowledge (page 103). The validity of this
value is tested with respect to the remaining limits.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or
the device is in the calibration mode.
CALCulate:LIMit:UPPer
DESCRIPTION
This specifies the upper limit when 2 comparator limits are selected.
SYNTAX
CALCulate:LIMit:UPPer <parameter>
PARAMETERS
The upper limit with an optional unit. Valid units: UOHM, MOHM,
OHM, KOHM, MAOHM. If no unit is specified, the unit of the value is
interpreted as OHM.
QUERY FORM
CALCulate:LIMit:UPPer?
RESPONSE
The upper limit of 2 comparator limits (value with unit).
BEMERKUNGEN
The value transferred with this command needs to be acknowledged
with CALCulate:LIMit:ACKNowledge (page 103).
The validity of this value is tested with respect to the remaining limits.
RESTRICTIONS
This command is not permissible if a measurement is in progress or
the device is in the calibration mode.