6.3.3 DISPlay Subsystem
DISPlay:CONTrast
DISPlay:RANGe
DISPlay:CURRent
DESCRIPTION:
The LCD contrast is adjusted here.
SYNTAX:
DISPlay:CONTrast <parameter>
PARAMETERS:
A value between 0 and 1.
1 -> maximum contrast,
0 -> minimum contrast
QUERY FORM:
DISPlay:CONTrast?
RESPONSE:
A value between 0 and 1.
RESTRICTION:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DESCRIPTION:
This activates/deactivates the measurement range display in the main menu.
SYNTAX:
DISPlay:RANGe <parameter>
PARAMETERS:
1 or ON -> The measurement range is displayed.
0 or OFF -> The measurement range is not displayed
QUERY FORM:
DISPlay:RANGe?
RESPONSE:
1 -> If the measurement range display is active.
0 -> If the measurement range display is inactive.
RESTRICTION:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
DESCRIPTION:
This activates/deactivates the measurement current display in the main menu.
SYNTAX:
DISPlay:CURRent <parameter>
PARAMETERS:
1 or ON -> The measurement current is displayed.
0 or OFF -> The measurement current is not displayed
QUERY FORM:
DISPLay:CURRent?
RESPONSE:
1 -> If the measurement current display is active.
0 -> If the measurement current display is inactive.
RESTRICTION:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.