16
of 148
9 pole Sub MinD
socket
View of socket
The 24-pole IEEE488 bus plug-connector is standardized and
has the displayed pin assignment.
To comply with applicable interference protection guidelines
(VDE 0871B), all interface cables and plugs must be shielded at
both ends.
Plug housing
:
Potential PE
Matching data cable
:
Type 5230-001
GND is internally grounded.
IEEE488 interface
PE-FE connection
In the case of ungrounded (or unearthed) test objects, PE-FE must
be connected to the short-circuit bridge.
In the case of grounded (or earthed) test objects, the short-circuit
bridge must be removed, otherwise double-ground compensation
currents might arise and lead to measurement errors.
Device socket
Black: FE
Functional earth
Yellow/green: PE
Protective earth
1 NC
2 TXD
3 RXD
4
5
Digital GND (internally grounded)
6
7 NC
8
9 NC
Plug housing
:
Potential PE
Counter-plug
:
Type 9900-V209
Matching data cable
:
Type 9305-K001
Bridged
inside the
device
RS232 interface
5 1
9 6