SENse:ENVelope:STATe
DESCRIPTION:
This activates/deactivates the max./min. function
SYNTAX:
SENSe:ENVelope:STATe <parameter>
PARAMETERS:
1 or ON -> The max./min. function is on.
0 or OFF -> The max./min. function is off.
QUERY FORM:
SENSe:ENVelope:STATe?
RESPONSE:
1 -> If the max./min. function is on.
0 -> If the max./min. function is off.
RESTRICTIONS:
This command is not permissible if the device is in the calibration mode.
SENSe:TCOMpensate:STATe
DESCRIPTION:
This activates/deactivates temperature compensation.
SYNTAX:
SENSe:TCOMpensate:STATe <parameter>
PARAMETERS:
1 or ON -> Temperature compensation is on.
0 or OFF -> Temperature compensation is off.
QUERY FORM:
SENSe:TCOMpensate:STATe?
RESPONSE:
1 -> If temperature compensation is on.
0 -> If temperature compensation is off.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
.
SENSe:ENVelope:CLEar
DESCRIPTION:
This deletes the minimum and maximum values.
SYNTAX:
SENSe:ENVelope:CLEar
PARAMETERS:
None
QUERY FORM:
No query form
RESPONSE:
—
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.