SENSe:TCOMpensate:TEMPerature:REFerence
DESCRIPTION:
This sets the reference temperature for temperature compensation.
SYNTAX:
SENSe:TCOMpensate:TEMPerature:REFerence <parameter>
PARAMETERS:
The reference temperature, optionally with a unit (C or CEL).
QUERY FORM:
SENSe:TCOMpensate:TEMPerature:REFerence?
RESPONSE:
The selected reference temperature and a corresponding unit (CEL).
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SENSe:TCOMpensate:TCOefficient
DESCRIPTION:
This selects one of 10 different temperature coefficients.
SYNTAX:
SENSe:TCOMpensate:TCOefficient <number>, <value>
PARAMETERS:
Number -> The number of the coefficient (1 to 10)
Value -> The value of the temperature coefficient.
(Value range: - 9999 to 9999)
QUERY FORM:
SENSe:TCOMpensate:TCOefficient? <number>
RESPONSE:
The temperature coefficient related to the number.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SENSe:FRESistance:RANGe:AUTO
DESCRIPTION:
This effects a switchover between automatic and manual range selection.
With the query form the present range will send devide with a comma.
SYNTAX:
SENSe:FRESistance:RANGe:AUTO <parameter>
PARAMETERS:
1 or ON -> Automatic range selection is active.
0 or OFF -> Manual range selection is active.
QUERY FORM:
SENSe:FRESistance:RANGe:AUTO?
RESPONSE:
1.2 OHM -> If automatic range selection and range 2
Ω
is active.
0.2 OHM -> If manual range selection and range 2
Ω
is active.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.