6.3.11 SENSe Subsystem
SENSe:TCOMpensate
DESCRIPTION:
This sets the type of temperature measurement for temperature
compensation.
SYNTAX:
SENSe:TCOMpensate <parameter>
PARAMETERS:
MAN -> The temperature value entered via the interface or
keypad is used.
Pt100 -> Measurement via the Pt100 input.
UINP -> Measurement via the external U-input.
IINP -> Measurement via the external I-input.
QUERY FORM:
SENSe:TCOMpensate?
RESPONSE:
MAN -> If manual temperature entry is selected.
Pt100 -> If temperature measurement via the Pt100 is selected.
UINP -> If temperature measurement via the ext. U-input is selected.
IINP -> If temperatu
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SENSe:FRESistance:REFerence
DESCRIPTION:
This select the reference value for
∆
% display.
SYNTAX:
SENSe:FRESistance:REFerence <parameter>
PARAMETERS:
The reference value, optionally with a unit. Valid units: UOHM, MOHM,
OHM, KOHM, MAOHM If no unit is specified, the unit of the value is
interpreted as OHM.
QUERY FORM:
SENSe:FRESistance:REFerence?
RESPONSE:
The selected reference value and unit.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.