111
of 148
SCALe:VOLTage
DESCRIPTION:
This scales the external U-input.
SYNTAX:
SCALe:VOLTage <u1>,<u2>,<t1>,<t2>
PARAMETERS:
u1: Current value 1 (related to t1)
u2: Current value 2 (related to t2)
t1: Temperature value 1 (related to u1)
t2: Temperature value 2 (related to u2)
Both pairs of values are used to calculate the linearization curve.
QUERY FORM:
SCALe:VOLTage?
RESPONSE:
u1,u2,t1,t2
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SCALe:CURRent
DESCRIPTION:
This scales the external I-input.
SYNTAX:
SCALe:CURRent <i1>,<i2>,<t1>,<t2>
PARAMETERS:
i1: Current value 1 (related to t1)
i2: Current value 2 (related to t2)
t1: Temperature value 1 (related to u1)
t2: Temperature value 2 (related to u2)
Both pairs of values are used to calculate the linearization curve.
QUERY FORM:
SCALe:CURRent?
RESPONSE:
i1,i2,t1,t2
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.