ACCess:COMPensation
DESCRIPTION:
This enables/disables access to the temperature-compensation menu.
SYNTAX:
ACCess:COMPensation <parameter>
PARAMETERS:
1 or ON -> Enables access to the temperature-compensation menu.
0 or OFF -> Disables access to the temperature-compensation
QUERY FORM:
ACCess:COMPensation?
RESPONSE:
1 -> If access to the temperature-compensation menu is enabled.
0 -> If access to the temperature-compensation menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
ACCess:COMMunicate
DESCRIPTION:
This enables/disables access to the interface menu.
SYNTAX:
ACCess:COMMunicate <parameter>
PARAMETERS:
1 or ON -> Enables access to the interface menu.
0 or OFF -> Disables access to the interface menu.
QUERY FORM:
ACCess:COMMunicate?
RESPONSE:
1 -> If access to the interface menu is enabled.
0 -> If access to the interface menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
ACCess:MEMory
DESCRIPTION:
This enables/disables access to the device-setting menu (for saving and
loading device settings).
SYNTAX:
ACCess:MEMory <parameter>
PARAMETERS:
1 or ON -> Enables access to the device-setting menu.
0 or OFF -> Disables access to the device-setting menu.
QUERY FORM:
ACCess:MEMory?
RESPONSE:
1 ->If access to the device-setting menu is enabled.
0 ->If access to the device-setting menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.