10. Appendix - Control Example
for device settings
Example for device setting:
Test object:
Coil approx. 1.5
Ω
(set Z)
Evaluation:
2 Limits
(set 2LI)
lower Limit 1.4
Ω
upper Limit 1.6
Ω
Temperatur compensation:
Material of the coil wire with Pt100 is copper (set TC 3980 ppm)
Meas. mode:
Continuous measurement
Meas. range:
2
Ω
manual range selection
Memory:
All the mentioned settings should be stored in the measure program 1
Device ON
:
Identification menu appears for 5 sec., after this time the device goes to
the main menu.
Now press following keys:
F4
:
Parameter selection menu appears, cursor pos. MEAS:PARAMETER
F1
:
Range select MAN/AUTO., with key
6
→
set MAN
(see manual pageSeite 37)
2
→
:
Meas. range
6
→
:
Press this key as much as "RANGE 2 OHM" appears.
2
→
:
Test object
6
→
:
Test object
:
set Z