BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual Interface I-300 Pro
85/102
u
Connect the interface to a LAN network. For connection details see Chapter 3.5
"Configuration", page 14
u
Navigate to the
Rotavapor Open Interface
sub-screen and select the option
[Yes]
, see Chapter 6.1 "Navigating the menu", page 44.
u
Confirm the action when asked.
ð
The interface shows the password
u
Restart the interface.
ð
The Rotavapor ® Open interface unit is set up.