Operation Manual Interface I-300 Pro
59/102
Parameters and readings
Equipment possibly required
Actual reading for current pressure in
distillation system
VacuBox (e.g. with Vacuum Pump
V-300/V-600)
Specified setting for maximum pump
speed
VacuBox (e.g. with Vacuum Pump
V-300/V-600)
Actual reading and specified setting
for rotation speed (evaporating flask)
Rotavapor R-300
Actual reading and specified setting
for heating bath temperature
Heating Bath B-301/B-305
Actual reading and specified setting
for cooling temperature
Recirculating Chiller F-3xx
Actual reading for vapor temperature
AutoDest sensor or vapor temperature
sensor
Actual reading for lift
NOTE
The pressure (vacuum) cannot be altered. The vacuum pump runs at the maximum
speed setting.
6.2.6
Executing AutoDest mode
Automatic distillation requires the use of an AutoDest sensor. The AutoDest sensor is
connected to the VacuBox (see Chapter 5.3 "Connecting AutoDest sensor to vapor
temperature sensor (optional accessory)", page 39) and measures the inlet and outlet
temperature of the coolant and the vapor temperature at the cooling condenser. From
those three readings, the system calculates the parameter settings required for opti-
mum distillation. During automatic distillation, the temperatures of the heating bath,
coolant and vapor are constantly measured and the specified setting for the pressure
adjusted accordingly.
CAUTION
Risk of personal injury and property damage from unexpected equip-
ment behavior
u
Always carefully check the pre-programmed settings before every distillation
process.
Navigation
➔
Operating modes
➔
AutoDest
A
Fig. 40: Selecting the option
[AutoDest]
on the
Operating modes
menu
Precondition:
R
An AutoDest sensor has been installed and calibrated, see Chapter 7.7 "Calibrat-
ing AutoDest sensor", page 80.