6 | Operation
BÜCHI Labortechnik AG
72/102
Operation Manual Interface I-300 Pro
u
To start a distillation process directly from the solvent library, press the function
button
START
.
ð
The distillation process is started directly.
u
To apply the settings for the selected solvent and return to the interface unit home
screen, press
OK
. On the home screen, the settings for the individual process pa-
rameters can be edited.
Fig. 58: Editing the specified settings for the selected solvent on the home screen
NOTE
The temperatures for heating bath and coolant should be set so as to be 40 °C apart.
6.5.2
Creating your own solvent library
You can also create your own solvent library in which you can enter details of up to
60 of your own solvents.
Navigation
➔
Tools
➔
Own solvent library
u
Navigate to the "Tools" screen and select the menu item "Own solvent library", see
Chapter 6.1 "Navigating the menu", page 44.