BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual Interface I-300 Pro
77/102
7
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec-
tion.
Any servicing and repair work which involves opening up the casing may only be car-
ried out by BUCHI service technicians.
u
Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
7.1
Cleaning the casing/display
u
Use a damp cloth to clean the device.
u
To clean the casing, only use ethanol or soapy water.
7.2
Performing a leak test
Use the leak test to check the system for leaks.
Navigation path
➜
Configuration
➜
Maintenance
➜
Leak test
Precondition:
R
All openings in the vacuum system are sealed.
u
Navigate to the
Maintenance
screen and select the menu item
[Leak test]
, see
Navigating the menu.
Fig. 65: Leak test screen
u
Press the function button START to start the leak test.
ð
The system is evacuated to 50 mbar within 2 minutes.
ð
After a stabilization period of 30 seconds, the leakage rate is shown.