BÜCHI Labortechnik AG
Product description | 3
Operation Manual Interface I-300 Pro
23/102
Action
Option
Explanation
[Record data]
On/Off
The following process data is recorded
on an SD card:
specified setting and actual readings for
pressure, rotation, heating bath, recircu-
lating chiller.
Actual vapor temperature.
Events such as AERATE and defoaming.
Difference between inlet and outlet tem-
perature at AutoDest sensor.
[Language]
Choice of language
used for control unit
display
English, Deutsch, Francais, Italiano, Es-
panol, Russian, Portugues, Japanese,
Chinese, Indonesian, Korean
[Button tone]
On/Off
A beep sounds when a function button or
the navigation control is pressed.
[Seal servicing in-
formation]
On/Off
The interface shows information about
regular servicing of the system seals.
The message appears periodically after
every 500 hours of rotation.
[Temperature unit]
Choice of unit for
indication of tem-
peratures
°C (Celsius), °F (Fahrenheit) or
K (Kelvin)
[Pressure Unit]
Choice of unit for
indication of (nega-
tive) pressure
hPa (hectopascals), mbar (millibars), torr
(= mmHg), mmHg (millimeters of mer-
cury)
[Height above sea
level]
Enter value
Altitude of location above mean sea
level: Max. 4000 m.
For determination of max. allowable
pressure when working with the solvent
library.
[Max. permissible
pressure]
Enter value
Max. pressure level present in the sys-
tem: max. 1300 mbar.
[Max. pump deliv-
ery]
Enter value
Max. pump speed in %: 10 –100 %.
[Display brightness]
Enter value
Display illumination level in %: 0 – 100
%.
[UTC offset]
Enter value
Manual entry of time difference between
local time and coordinated universal time
(UTC).
[Current date]
Enter value
Manual entry of date. Necessary for
identification of recorded data.
[Current time]
View
Display of local time. Can be changed by
means of the action
[UTC offset]
.
Necessary for identification of recorded
data.
[eco mode]
On/Off and entry of
figures
See Chapter 6.6 "Activating eco mode",
page 73