BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual Interface I-300 Pro
81/102
u
Wait until the difference indicated has stabilized.
ð
The two condenser sensors are now at the same temperature.
Navigation path
➜
Configuration
➜
Maintenance
➜
Calibrate AutoDest sensor
u
Navigate to the
Maintenance
sub-screen and select the menu item
[AutoDest
sensor]
, see Navigating the menu.
u
Press the navigation control to choose OK and confirm the calibration offset.
7.8
Calibrating the pressure sensor
The pressure sensor is calibrated at the factory by BUCHI prior to delivery. However,
it can be recalibrated with the aid of an external reference pressure gauge at any
time.
Navigation path
➜
Configuration
➜
Servicing
u
Navigate to the "Configuration" screen and select the menu item "Servicing", see
Chapter 6.1 "Navigating the menu", page 44.
Fig. 68: Selecting item on the "Service" screen