3 | Product description
BÜCHI Labortechnik AG
22/102
Operation Manual Interface I-300 Pro
Action
Option
Explanation
[Test Results]
Info
Shows the leak tests carried out. See
Chapter 7.3 "Opening leak test history",
page 78
[Leak test]
START/Information
For carrying out a leak test on the distil-
lation system . See Chapter 7.2 "Per-
forming a leak test", page 77
[Reset duty hours]
Info
Hours of rotation since last service.
Resets the hours of rotation recorded.
Settings submenu
The submenu
[Settings]
contains options for settings that can be made on the distilla-
tion system.
Action
Option
Explanation
[Mobile connection
QR code]
View
Interface unit shows QR code for the
connected distillation system.
[Mobile connection
password]
View
For viewing password and entering on
mobile device (alternative to QR code)
[On finish: Aerate
system]
On/Off
System is vented after automatic or man-
ual termination of distillation.
[On start: start rota-
tion]
On/Off
Evaporating flask starts rotating when
distillation is started.
[On finish: Stop ro-
tation]
On/Off
Evaporating flask stops rotating after au-
tomatic or manual termination of distilla-
tion.
[On start: immerse
flask]
On/Off
Evaporating flask is automatically im-
mersed in the heating bath when distilla-
tion is started.
[On finish: lift out
flask]
On/Off
Evaporating flask is automatically lifted
out of the heating bath after automatic or
manual termination of distillation.
[On finish: stop
heating]
On/Off
Heating of heating bath stops after auto-
matic or manual termination of distilla-
tion.
[On finish: stop
cooling]
On/Off
The Recirculating Chiller F-3xx automati-
cally switches off after automatic or man-
ual termination of distillation.
[On finish: play
sound]
On/Off
An audible signal is sounded after auto-
matic or manual termination of distilla-
tion.
[Pressure hystere-
sis]
Entry of pressure
Entry of figure for the maximum allow-
able difference between the actual vac-
uum and the specified figure before the
vacuum pump switches on again.