BÜCHI Labortechnik AG
Product description | 3
Operation Manual Interface I-300 Pro
25/102
Action
Option
Explanation
[Pressure calibra-
tion]
Calibration of pres-
sure sensor and
entry of reference
pressure
The pressure sensor is calibrated in five
stages for the following set pressures:
- Approx. 950 mbar (ambient pressure)
- 800mbar
- 600mbar
- 400mbar
- 200mbar
- 10mbar
See Chapter 7.8.2 "Simple calibration",
page 83.
[Load factory cali-
bration]
Calibration reset
The current pressure sensor calibration
data is overwritten by the factory calibra-
tion settings. See Chapter 7.8.3 "Loading
factory calibration", page 84.
System Information submenu
The submenu
[System information]
contains details of the currently connected labora-
tory equipment and information on network connection diagnosis.
3.6.5
Libraries menu
Action
Option
Explanation
[Solvent library]
Select
The solvent library item-
izes the commonly used
solvents in alphabetical or-
der.
[Own solvent library]
For creating your own sol-
vent library
See Chapter 6.5.2 "Creat-
ing your own solvent li-
brary", page 72
[Startup info]
View
Shows the home screen.
3.6.6
Symbols on the status bar
Symbol
Status
Manual distillation
Distillation using a method
Drying
A
AutoDest distillation
Distillation running
SD
The memory card has been inserted.
Continuous pumping
The instrument is connected to the BUCHI Cloud.
Timed distillation