6 | Operation
BÜCHI Labortechnik AG
60/102
Operation Manual Interface I-300 Pro
R
The system is not leaking. See Chapter 7.2 "Performing a leak test", page 77
u
Navigate to the
Operating modes
menu and select the menu item
[AutoDest]
,
see Chapter 6.1 "Navigating the menu", page 44.
ð
The display shows the home screen with the symbol for automatic distillation
in the status bar.
Fig. 41: Home screen showing
[AutoDest]
symbol
u
Select and edit the required settings for individual process parameters as neces-
sary. When doing so, make sure that the temperatures for heating bath and
coolant are set so as to be 40 °C apart.
In this operating mode, the following parameters and readings are available:
Parameters and readings
Equipment possibly required
Specified setting for vacuum
Actual reading for current pressure in
distillation system
VacuBox (e.g. with Vacuum Pump
V-300/V-600)
Actual reading and specified setting
for rotation speed
Rotavapor R-300
Actual reading and specified setting
for heating bath temperature
Heating Bath B-301/B-305
Actual reading and specified setting
for cooling temperature
Recirculating Chiller F-3xx
Actual reading for vapor temperature
AutoDest sensor
Actual reading for lift
Starting automatic distillation
Precondition:
R
Process parameters have been set.
u
Press the function button START.