BÜCHI Labortechnik AG
Product description | 3
Operation Manual Interface I-300 Pro
17/102
3.5.4
LegacyBox (connections)
Connection of a LegacyBox is required whenever legacy BUCHI laboratory equip-
ment is to be controlled via the Interface I-300 Pro. The LegacyBox is connected to
the distillation system using a standard BUCHI communication cable and has other
connection options such as an RS-485 communication port. There is also the option
of incorporating pumps of other makes in the Rotavapor system and controlling them
via the interface. In that case, the pump requires a 0 – 10 V input.
The LegacyBox is fixed to a Rotavapor R-300 or a laboratory stand clamp.
2
4
3
1
6
5
Fig. 4: Connections on the LegacyBox
1
Standard BUCHI communication port
(COM)
4
RS-232 (not used)
2
0 – 10 V communication port – for
connecting pumps of other makes
5
External power supply – for genuine
BUCHI mains adaptor, 30 V, 30 W
(see Accessories)
3
RS-485 communication port – for
connecting legacy BUCHI laboratory
equipment (Vacuum Pumps V-700 /
V-710, Rotavapors: R-210 / R-215,
Recirculating Chiller F-1xx)
6
On/Off switch
NOTE
The external power supply is only necessary if the LegacyBox is not connected to a
Rotavapor or a vacuum pump and is used for other vacuum control applications in-
stead. Otherwise the vacuum pump or the Rotavapor supplies the LegacyBox with
power.
0 – 10 V communication port
This connection comprises a 0 – 10 V output for connecting pumps of other makes
and electrically isolated relay contacts. There is also a valve control contact.