BÜCHI Labortechnik AG
Operation | 6
Operation Manual Interface I-300 Pro
51/102
Navigation path
➔
Operating modes
➔
Manual
Fig. 27: Selecting the option
[Manual]
on the
Operating modes
menu
u
Navigate to the
Operating modes
menu and select the menu item
[Manual]
, see
Chapter 6.1 "Navigating the menu", page 44.
ð
The display shows the home screen with the symbol for manual distillation in
the status bar.
Fig. 28: Home screen showing
[Manual]
symbol
u
Enter the required settings for the process parameters.
In this operating mode, the following parameters and readings are available:
Parameters and readings
Equipment possibly required
Specified setting for vacuum
Actual reading for current pressure in
distillation system
VacuBox (e.g. with Vacuum Pump
V-300/V-600)
Actual reading and specified setting
for rotation speed (evaporating flask)
Rotavapor R-300
Actual reading and specified setting
for heating bath temperature
Heating Bath B-301/B-305
Actual reading and specified setting
for cooling temperature
Recirculating Chiller F-3xx
Actual reading for vapor temperature
AutoDest sensor or vapor temperature
sensor
Actual reading for lift