82 Lumax Technical Manual
2.8.3.6 Third Programmable Shock Pathway
Lumax 500/540 Models only
The Lumax 500/540 models offer three programmable shock
delivery configurations. The defibrillation shock pathways include
a “non-active housing” shock path and a shock paths between the
ICD/CRT-D housing and either or both shock coils when
connected to a dual coil shock lead.
The Shock Configuration parameter operates independently of
Shock Polarity. So each of the pathways described below can be
reversed for delivered defibrillation/cardioversion shocks.
Table 24
outlines the current shock pathways that are available
with Lumax 500/540 models.
Table 24 Lumax 500/540 Programmable Shock Pathways
#
Electrical Pathway
RV
Î
SVC and Housing (can)*
RV
Î
SVC
RV
Î
Housing (can)
- 300/340 Models are limited to this configuration.
2.8.4 Progressive Course of Therapy
By design, the Lumax ICDs/CRT-Ds will deliver more aggressive
therapy for each successive attempt within a single detected
episode. Therefore, the device will not deliver ATP1 therapy
following ATP2 therapy, and will not deliver ATP therapy following
a high voltage defibrillation shock.
When Progressive Course of Therapy is turned ON, the ICD/CRT-
D will always deliver a maximum energy shock after re-detecting
in an arrhythmia class with programmed shock energy less than
or equal to the previously delivered therapy. In addition, the
ICD/CRT-D blocks all ATP therapy during the current episode if a
shock has already been delivered during the episode.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......