Lumax Technical Manual 37
Parameter
Range
Standard
S1-S2 interval
200...(10)...500 ms
(absolute);
70...(5)...95 % (adaptive)
[70 %]
Minimal ATP interval 200...(5)...300 ms
[200 ms]
ATP timeout
OFF,
00:15...(00:15)...05:00
mm:ss
OFF
ATP optimization
OFF, ON
OFF
ATP pulse amplitude 7.5 V
7.5 V
ATP pulse width
1.5 ms
1.5 ms
ATP One-Shot Parameter (ATP in VF)
ATP type
OFF, Burst, Ramp, Burst +
PES
OFF
S1 number
1...(1)...10
[5]
R-S1 interval
200...(10)...350 ms
(absolute);
70...(5)...95 % (adaptive)
[80%]
S1 decrement
5...(5)...40 ms
[10ms]
S1-S2 interval
200...(10)...350 ms
(absolute);
70...(5)...95 % (adaptive)
[70%]
Stability
12%
12%
ATP attempts
1
[1]
ATP pulse amplitude 7.5 V
7.5 V
ATP pulse width
1.5 ms
1.5 ms
Post-Shock Pacing
Mode
DDI at permanent
DDD(R), DDI (R), AAI(R)
VDI at permanent
VDD(R), VDI(R),
VVI at permanent VVI(R),
OFF
Basic rate
30 .. (5)..100..(10) .. 160
ppm
60 ppm
Rate hysteresis
OFF, -5 .. (-5) .. –65 bpm
OFF
AV delay
50..(10)..350 ms (fixed AV
delay)
140 ms
Post-shock duration
OFF, 00:10 .. (00:10) ..
00:50, 01:00 .. (01:00) ..
10:00 mm:ss
00:10 mm:ss
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......