Lumax Technical Manual 93
2.9.8 Mode Switching
DR and HF versions only
Mode switching is designed to avoid tracking of atrial arrhythmias.
In the presence of a high atrial rate, the bradycardia pacing mode
is automatically reprogrammed to a non-atrial tracking mode. The
modes available during mode switching are as shown in
Table 27
. Mode switching is not available during the post-shock
pacing period. Mode Switching is only available with the Lumax
DR and Lumax HF ICDs/CRT-Ds.
Table 27: Mode Switching Modes
Programmed Mode
Converted Mode
DDDR
DDIR
DDI
DDD
DDIR
DDI
VDDR
VDIR
VDI
VDD
VDIR
VDI
Mode switching is initiated in atrial tracking modes when the atrial
rate, defined by the programmable mode switch
Intervention
Rate
is achieved. However, mode switching will not occur until
the Mode Switch
Activation
Criteria
is also met. The activation
criterion is a programmable
X out of 8
high rate intervals as
programmed. RVsense Triggering can be programmed separately
during mode switching.
After switching to a non-atrial tracking mode, the ICD/CRT-D
activates a
Y out of 8
counter that deactivates mode switching
when Y number of cardiac cycles out of the last 8 are below the
Intervention Rate. When this
Deactivation
Criteria
parameter is
fulfilled, the ICD/CRT-D returns to the normal programmed pacing
mode.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......