30 Lumax Technical Manual
6. Technical Specifications
The following are the technical specifications for the
Lumax ICDs/CRT-Ds. The ranges are presented in the format:
x…(y)…z
where x = the lowest value, y = the increment, and z = the largest
value.
Mechanical Properties
Lumax
300/500
Lumax
340/540
Lumax
500 HF-T
Lumax
340 HF /
540 HF-T
Dimensions
66 mm x
55 mm x
12 mm
66 mm x
55 mm x
13 mm
66 mm x
59 mm x
12 mm
66 mm x
59 mm x
13 mm
Volume
34.6 cm
3
37.2 cm
3
37.1 cm
3
39.8 cm
3
Mass
81 g
92 g
83 g
94 g
Housing
Titanium
Header
Epoxy resin
Seal Plug
Silicone
VR-T
Lead Ports
1 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
DR-T
Lead Ports
2 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
3 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
Parameter
Range
Standard
Bradycardia
Atrial Sensing Parameters
Sensing
STD - standard,
OFF - inactive
STD
Minimum threshold
0.2...(0.1)...2.0 mV
0.4 mV
Far-field protection
after Vp
50...(25)...225 ms
75 ms
Far-field protection
after Vs
Off, 25...(25)...225 ms
75 ms
Upper threshold
50; 75; 87.5 %
50 %
Lower threshold
12.5; 25; 50 %
25 %
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......