Lumax Technical Manual 63
LV T-wave Protection
- Used to eliminate unintended pacing in
the vulnerable period of the left ventricle. This feature is only
used when left ventricular sensing is active.
HF versions only.
2.3 Automatic Threshold Measurement
(ATM)
Lumax 500/540 Models only
The Lumax 500/540 models have automatic threshold
measurement (ATM) feature for determining ventricular pacing
thresholds. This feature is separately programmable for the right
(RV) and left (LV) ventricles.
ATM is initiated once each day, typically during the nighttime while
the patient is normally sleeping. It can measure the right and the
left ventricular pacing thresholds and stores this information for
use in trends of daily threshold values. This information is
available on the programmer during in-office follow-ups and via
BIOTRONIK’s Home Monitoring system. The permanent pacing
amplitude is not adjusted by the ATM feature.
2.3.1 Functional Description
The ATM features of Lumax 500/540 are based on the evaluation
of the morphology of the ventricular evoked response (VER) to
determine if the pacing pulse has captured the myocardium. In
order to measure the pacing threshold, a sequence of two
algorithmic steps is carried out:
•
Signal Quality Check (SQC)
•
Threshold
Measurement
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......