Lumax Technical Manual 45
Anomaly
Possible Effect on Patient
or Implant Procedure
Lumax Application (Programmer Software)
Dual Chamber devices only:
When programmed to DDDR
mode and with sensor activity it
is possible for a sudden rate
drop at start of a capacitor
reformation procedure. After
the reformation the pacing rate
again sensor indicated rate.
ID# 19607
Limited effect on patient,
temporary rate drop to
programmed post-shock
pacing rate ensures sufficient
support.
An ineffective pacing program
may continue during the
threshold test at programming
distances close to the loss of
communication. Lifting the
wand corrects the problem.
ID# 17903
Limited potential effect on
patient, non-capture is quickly
recognized and removal of
programming wand is typical
response. This anomaly can
only occur during manual
threshold testing (in
physician’s office).