44 Lumax Technical Manual
Anomaly
Possible Effect on Patient
or Implant Procedure
Lumax Application (Programmer Software)
SVT Detection Counters are
not incremented when SMART
is OFF and SVTs are detected,
still they are correctly recorded
in the Holter episodes.
ID# 21702
Limited effect on patient,
display of incorrect/missing
diagnostic data may cause
temporary user confusion
In non-rate-adaptive pacing
modes sensor parameters are
not accessible to allow their
adjustment in order to optimize
sensor statistics. The default
settings may not be
appropriate for optimal sensor
statistics settings. ID# 21233
Limited effect on patient,
sensor data collected from
non-rated-adaptive pacing will
not be optimized for later
change to rate-adaptive
pacing settings.
In Sensing Test, Freeze
Window Display of P/R wave
amplitude annotations may not
match caliper measurement
value. Freeze Window displays
filtered IEGM signal (caliper
value) whereas annotation
results from calculations using
on unfiltered IEGM. ID# 20899
Limited effect on patient,
display/printout of
incorrect/missing diagnostic
information may cause
temporary user confusion
Atrial impedance
measurements will not be
performed when intrinsic atrial
rhythm is above 110 bpm, but
the user does not receive any
response other than the
missing measurement value.
ID# 20475
Limited effect on patient,
display/printout of missing
diagnostic information may
cause temporary user
confusion
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......