Lumax Technical Manual 35
Parameter
Range
Standard
Onset
Onset in VT1/2 with
SMART
20%
[20%]
Onset VT1 without
SMART
OFF; 4…(4)…32%
20%
Onset VT2 without
SMART
OFF; 4…(4)…32%
20%
Stability
Stability in VT1/2
with SMART
12%
[12%]
Stability VT1 without
SMART
OFF; 8…(4)…48 ms
24 ms
Stability VT2 without
SMART
OFF; 8…(4)…48 ms
24 ms
SMART Detection
SMART detection
VT1
OFF, ON
[ON]
SMART detection
VT2
OFF, ON
[ON]
Sustained VT (without SMART Detection and without
SMART Redetection)
Sustained VT
OFF, 00:30, 01:00, 02:00,
03:00, 05:00, 10:00,
15:00, 20:00, 25:00, 30:00
[mm:ss]
[OFF]
Forced Termination (with SMART Detection Including
SMART Redetection)
Forced termination
OFF; 1…(1)…15 min
[1 min]
Redetection Counter
Redetection counter
VT1
10...(2)...30
[12]
Redetection counter
VT2
10...(2)...30
[10]
SMART Redetection
SMART redetection
(VT1 & VT2)
OFF, ON
[ON]