34 Lumax Technical Manual
Parameter
Range
Standard
Rate drop
0.25..(0.25)..1.25
ppm/cycle
0.5
ppm/cycle
Upper Tracking Rate (UTR)
Upper tracking rate
90..(10)..160 ppm
130 ppm
Upper tracking rate
atrium
OFF, 240 ppm
240 ppm
Mode Switching
Intervention rate
OFF, 100..(10)..250 ppm
160 ppm
Activation criterion X
3..(1)..8
5
Deactivation criterion
Z
3..(1)..8
5
Mode
DDI, DDIR at permanent
DDD(R)
VDI, VDIR at permanent
VDD(R)
DDI
[VDI]
Change in basic rate OFF, +5 ... (5) ...+30 ppm
+10 ppm
Post-Mode Switch Response (PMSR)
Post-ModeSw rate
OFF, +5 ... (5) ...+50 ppm
+10 ppm
Post-ModeSw
duration
1..(1)..30 min
1 min
PMT Protection
PMT detection /
termination
OFF, ON
ON
VA criterion
250..(10)..500 ms
350 ms
Detection
Detection / Therapy
ON, OFF
ON
Interval
Interval VT1
OFF, 270...(10)...600 ms
OFF
Interval VT2
OFF, 270...(10)...500 ms
OFF
Interval VF
OFF, 200...(10)...400 ms
300 ms
Detection Counter
Detection counter
VT1
10...(2)...60
[16]
Detection counter
VT2
10...(2)...40
[14]
Detection counter VF
– X
6...(1)...30
8
Detection counter VF
– Y
8...(1)...31
12
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......