Lumax Technical Manual 55
Table 18. Lead Configurations
Configuration
Explanation
RV
Only
Sensing takes place between
the tip and ring electrodes of
the right ventricular lead.
Sensing
LV
Only
Sensing takes place between
the tip and ring electrodes
(bipolar) or the tip electrode of
the left ventricular lead and the
CRT-D housing (unipolar).
Pacing
RV & LV
Together (BiV)
Pacing configuration is
programmable between the tip
and ring electrodes of the right
and left ventricular leads.
See Figure 3
Pacing
RV
Only
Pacing takes place between
the tip and ring electrodes of
the right ventricular lead.
Figure 3 Programmable BiV Pacing Configurations