
102 Lumax Technical Manual
To permit evaluation of the sensing function, the pacing rate must
be lower than the patient's intrinsic rate. In demand pacing, the
proper sensing function can be recognized if the interval between
intrinsic events and the following pacing pulse equals the basic
interval (if no Hysteresis is programmed). The following
parameters are programmable when performing the
measurements:
•
Pacing
Mode
•
Pacing
Rate
•
Upper Rate (Dual Chamber modes only)
•
Pulse
Amplitude
•
Pulse
Width
For evaluation of the sensing function, the pulse generator
features an intracardiac electrogram (IEGM) with marker signals
to indicate sensed and paced events.
2.10.2 Pacing Impedance Measurements
The Lumax ICDs/CRT-Ds have the ability to perform automatic
and manual pacing impedance measurements. The devices can
measure the pace impedance in any of the pacing configurations;
RA, RV, LV or BiV. In addition, the LV pacing polarity is also
programmable for impedance measurements (See Section 2.1).
The impedance is measured using a triggered mode with paces
of 2.4 Volts and 0.5 ms, or programmed value (whichever is
higher). During automatic measurements both atrial (only with an
atrial pacing mode) and ventricular (RV, LV or BiV) impedances
are measured
.
However, they will not be measured if the normal
pacing output is programmed to a value greater than the
impedance test output (2.8 V).
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......