Lumax Technical Manual 29
5.3 Explantation
Explanted ICDs/CRT-Ds, lead systems, and accessories may not
be reused. Please complete the appropriate out of service (OOS)
form and return it to BIOTRONIK with the explanted devices. All
explanted devices should be sent either to the local BIOTRONIK
representative or the BIOTRONIK home office for expert disposal.
Contact BIOTRONIK if you need assistance with returning
explanted devices. If possible, the explanted devices should be
cleaned with a sodium-hyperchlorine solution of at least 1%
chlorine and then washed with water prior to shipping.
The pulse generator should be explanted before the cremation of
a deceased patient.
W
ARNING
Unwanted Shocks
– Always program ICD Therapy to OFF
prior to handling the device to prevent the delivery of serious
shocks to the patient or the person handling the device
during the implant procedure.
C
AUTION
Device Incineration
– Never incinerate the ICDs/CRT-Ds
due to the potential for explosion. The ICD/CRT-D must be
explanted prior to cremation.
Explanted Devices
– Return all explanted devices to
BIOTRONIK.