Lumax Technical Manual 27
Table 31 Lumax VR-T Longevity Estimates
Pacing Support
Device
Shocks per
year
0%
15%
50%
100%
4
8.6
8.4
7.8
7.2
8
7.3
7.1
6.8
6.3
500 VR-T
12
6.3
6.2
5.9
5.6
4
9.0
8.8
8.3
7.7
8
7.3
7.2
6.9
6.4
540 VR-T
12
6.2
6.1
5.8
5.5
The dual-chamber Lumax 500/540 variants (Lumax 500 DR-T,
and Lumax 540 DR-T) of ICDs are intended to operate for more
than 6 years under normal use.
Table 32
provides longevity
estimates for these devices. The table provides several different
support scenarios. It is assumed that the shocks are equally
spaced throughout the life of the ICD.
Table 32 Lumax DR-T Longevity Estimates
Pacing Support
Device
Shocks per
year
0%
15%
50%
100%
4
8.2
7.8
7.0
6.0
8
7.0
6.7
6.1
5.4
500 DR-T
12
6.1
5.9
5.4
4.8
4
8.6
8.2
7.4
6.5
8
7.1
6.8
6.2
5.6
540 DR-T
12
6.0
5.8
5.4
4.9
The Lumax 500/540 HF-T variants of ICDs are intended to
operate for more than 6 years under normal use.
Table 33
provides longevity estimates for these devices with Greatbatch
batteries. The table provides several different support scenarios.
It is assumed that the shocks are equally spaced throughout the
life of the ICD.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......