Lumax Technical Manual 7
1.4.1 Sterilization, Storage, and Handling
Device Packaging
- Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because the
integrity of the sterile packaging may be compromised. Return
the device to BIOTRONIK.
Re-sterilization
- Do not re-sterilize and re-implant explanted
devices.
Storage (temperature)
- Store the device between 5° to 45°C
(41° - 113° F) because temperatures outside this range could
damage the device.
Storage (magnets)
- To avoid damage to the device, store the
device in a clean area, away from magnets, kits containing
magnets, and sources of electromagnetic interference (EMI).
Temperature Stabilization
- Allow the device to reach room
temperature before programming or implanting the device
because temperature extremes may affect initial device function.
Use Before Date
- Do not implant the device after the USE
BEFORE DATE because the device may have reduced longevity.
1.4.2 Device Implantation and Programming
Blind Plug
- A blind plug must be inserted and firmly connected
into any unused header port to prevent chronic fluid influx and
possible shunting of high energy therapy.
Capacitor Reformation
- Infrequent charging of the high voltage
capacitors may extend the charge times of the ICD/CRT-D. The
capacitors are reformed automatically at least every 85 days. For
further information, please refer to
Section 2.11.4
, Capacitor
Reforming.
Connector Compatibility
– ICD/CRT-D and lead system
compatibility should be confirmed prior to the implant procedure.
Consult your BIOTRONIK representative regarding lead/pulse
generator compatibility prior to the implantation of an ICD/CRT-D
system. For further information, please refer to
Appendix A
.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......