Lumax Technical Manual 73
2.8.1 Therapy Options
The Lumax ICDs/CRT-Ds offer multiple therapy options for each
tachyarrhythmia class (VT1, VT2, VF). Therapy options (up to 10
ATP sequences and 8 shocks) are available for the VT1 and VT2
zones, whereas ATP One Shot and up to 8 shock therapies are
available for the VF class. The specific characteristics of an ATP
and shock therapy are independently programmed for each
VT/VF zone.
The ATP and shock therapy options are discussed in detail in the
following sections.
2.8.2 Anti-Tachycardia Pacing (ATP)
Anti-tachycardia pacing therapy (ATP) is available in both VT
detection zones. Available modes of ATP include
Burst
,
Ramp
,
and
Burst + PES
(Programmed Extra Stimuli). In addition, the
Burst and Ramp modes allow interval scanning of the
R-S1
Interval
, the
S1 Decrement
, or both. The
Attempts
parameter
determines the number of burst schemes to be delivered before
the scan parameter is incremented.
Burst
– This mode will deliver a series of pacing stimuli with user
defined duration of the burst (
Number S1
), coupling cycle length
(
R-S1
) and burst rate (
S1-S1
). The coupling interval and the start
interval are calculated from the intrinsic R-R average.
Ramp
- This mode will deliver a series of pacing stimuli with the
above options including a parameter which decrements each
successive stimuli interval in the burst.
Burst + PES
- This mode provides a pulse train followed by one
additional timed stimuli. The coupling cycle length of the burst
and each extra stimulus (
S1-S2 Interval
) is individually
programmed either as an adaptive value (as a percentage) or as
an absolute value (expressed in milliseconds).
Ventricular Pacing
- This parameter defines the type of
ventricular pacing performed during delivery of ATP sequences in
CRT-Ds and is programmable to biventricular (VV delay = 0 ms)
or right ventricular only.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......