32 Lumax Technical Manual
Parameter
Range
Standard
Shock Path (Valid for
all shocks including the
pain-free shock
impedance)
RV -> SVC + ICD (Housing)
RV -> ICD (500/540 only)
RV -> SVC (500/540 only)
RV -> SVC +
ICD
Pulse Amplitudes and Pulse Widths
Pulse amplitude
0.2...(0.1)...6.2, 7.5 V
2.8 V
Pulse width
0.4, 0.5, 0.7, 1.0, 1.2, 1.5
ms
0.4 ms
Automatic Threshold Measurement (ATM)
For Lumax 500/540 Models Only
RV
ON; OFF
ON
1
LV (HF-T Only)
ON; OFF
ON
Mode
Modes
DDD, DDDR, DDI, DDIR,
VDD, VDDR, VDI, VDIR,
VVI, VVIR, AAI, AAIR,
OFF
DDD
Basic Rate Day/Night
Basic rate
30..(5)..100..(10)..160
ppm
60 ppm
Night rate
OFF, 30..(5)..100 ppm
OFF
Night beginning
00:00..(1 min)..23:59 h:m
[22:00 h:m]
Night ending
00:00..(1 min)..23:59 h:m
[06:00 h:m]
Rate Hysteresis
Rate hysteresis
OFF, -5..(-5)..-90 ppm
OFF
Repetitive
OFF; 1..(1)..15
[OFF]
Scan
OFF; 1..(1)..15
[OFF]
AV Delay
AV delay
Low, medium, high, fixed,
individ.
Low
AV delay 1
40..(5)..350 ms
150 ms
AV rate 1
30..(10)..120 ppm
60 ppm
AV delay 2
40..(5)..350 ms
120 ms
AV rate 2
70..(10)..160 ppm
130 ppm
Sense compensation OFF; -5..(-5)..-60 ms
-30 ms
1
The Standard Program for the ATM feature is OFF only for the
Lumax 500/540 VR-T
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......