42 Lumax Technical Manual
Table 13: Percentage of Ventricular Pacing – 3-Month and
6-Month Follow-ups
Percentage of
Ventricular
Pacing
3-Months
Patients
(percentage)
6-Months Patients
(percentage)
<80%
9 (7.4%)
4 (4.0%)
81 – 85 %
4 (3.3%)
2 (2.0%)
86 – 90 %
13 (10.7%)
9 (9.1%)
91 – 95 %
19 (15.7%)
20 (20.2%)
96 – 100 %
76 (62.8%)
64 (64.7%)
Totals
121 (100%)
99 (100%)
The majority of the follow-ups (84.9%) show a percentage of
ventricular pacing of 91% or more at Six-Months.
4. The purpose of secondary endpoint 4 is to evaluate
improvement in functional capacity as measured by the six
minute walk test. The six minute walk test is a well-accepted
measure of functional capacity and exercise tolerance. Also,
this test more closely mimics the patient’s day-to-day
activities than maximal exercise testing.
Table 14
summarizes the six minute walk test distance at
Baseline and the Six-Month follow-up for patients in the study
group and the control group.