2 Lumax Technical Manual
2.11.9.1 Event Counters
The total percentages of atrial sensed (PVARP and atrial
refractory period), atrial paced, right ventricular sense (VES and
refractory period) and ventricular paced events since the statistics
package was initiated are available. The total percentage of time
for each of the above listed events is also available. The total
numbers of atrial and ventricular events are also recorded.
2.11.9.2 Event Episodes
The total percentages of several timing events are displayed
under the Event Episode heading. These include:
Pacing Counters (Brady)
•
AV-Sequences
−
Intrinsic Rhythm (AsVs)
−
Conducted Rhythm (AsVp)
−
Atrial Paced Rhythm (ApVs)
−
Complete Paced Rhythm (ApVp)
Pacing Counters (CRT)
•
LV-RV-Sequences
−
BiVp (RVp-LVp or LVp-RVp
−
RVs-LVp
−
RVp-noLVp, RVs-noLVp (LV T-wave protection ON)
−
LVp, noLVp (LV pacing only)
•
VES-Triggered
Resynchronization
−
VES-LVp (Triggering: RVs+RVES)
−
VES-noLVp (LV T-wave protection ON)
−
VES per hour
2.11.9.3 Rate Trends
The device counts the number of paced and sensed events and
displays them in two different graphs (rate and paced). These
trends are available for 24 hour periods and as long-term
durations.